Title Mold Air-O-Cell Sampling Guide

AIR-O-CELL SAMPLING GUIDE                                Air-O-Cell Cassettes $199 (50 Pack) 
                                                                                       BUY NOW !               1-888-958-8170
 

PROCEDURE FOR AIR-O-CELL SAMPLING

The Air-O-Cell™ Air Sampling cassette is a sampling device designed for the rapid collection and analysis of a wide range of airborne aerosols. These include fungal spores, pollen, insect parts, skin cell fragments, fibers, and inorganic particulates. Air enters the cassette, the particles become impacted on the sampling substrate, and the air leaves through the exit orifice. The airflow and patented cassette housing is designed in such a way that the particles are distributed and deposited equally on a special glass slide contained in the cassette housing called the ``trace."

BENEFITS
Useful for initial site testing, especially if fungal growth is not visible. Quick and simple procedure. Fast turn around times available. Low chance of sample contamination.

DISADVANTAGES
Fungi cannot be fully speciated with this method. For example, Aspergillus sp. and Penicillium sp. are normally reported together due to the similarities in spore morphology. Spore viability cannot be assessed.

MATERIALS
Air-O-Cell™ cassette Vacuum pump suitable for operating at 15 liters per minute (lpm) Rotameter (for calibration of vacuum pump) Flexible tubing Recommended Sampling Procedure The cassette is designed to operate at a recommended flow rate of 15 lpm. Lower flow rates may result in a loss of some spores and the accumulation of others in a non-uniform manner. Higher flow rates may damage the spores. Therefore, it is important not to run the sampling pumps at lower flow rates for longer times to achieve the recommended air volume. Calibration of Sampling Flow Rate A specially designed bubble tube, available from the manufacturer, can be used to set the flow rate for the ½" Air-O-Cell mini pump. It can not be used with any other style of pump or Mini pump and any other type of sampling pump.

SAMPLING DURATION
The sampling time is dependent on the density of particulate in the environment. It is important not to overload the sample, otherwise it will be impossible to assess the types of spores, pollen and particulates that are present. The following table represents typical sample times to attain a sharply defined trace with good dispersion of the spores: Recommended Sampling Intervals for the Air-O-Cell™ Air Sampling Cassette at Typical Collection Flow Rates Typical Environmental Conditions Flow Rates Collection at 15 lpm Clean ``office" or outdoors (no visible dust) 10 minutes Indoor environment, high activity & personnel 5 minutes Indoor environment, drywall renovation or heavy industrial dust 1 minute

SAMPLING PROCEDURE
1.Prior to sampling, calibrate the pump to 15 liters per minute.

2.Remove and retain tape seal covering Air-O-Cell™ inlet and outlet.

3.Attach the outlet (round hole) to the supplied tubing adapter (available upon request), or use standard ½" PVC tubing (for use with high volume pumps only).

4.Start the sampling pump, and sample for an appropriate period of time (see chart above).

5.Remove Air-O-Cell™ from tubing, and reseal with the original tape. Label sample.

6.Complete an EMSL Chain of Custody (COC), available on the websitedetailing client name and information, project name or number, sample #, description of area, and volume of air used.

7.To reduce shipping damage, it is recommended that the Air-O-Cell™ be placed in a corrugated box to ensure safe arrival at the laboratory.

QUALITY CONTROL SUGGESTIONS
1. An effective interpretation is based on the comparison of indoor and outdoor samples. An outdoor sample will help demonstrate whether spore amplification is occurring indoors.

2. Obtain a control sample from a non-complaint area for comparison.

3. Sending a blank cassette for analysis periodically is a good practice.

4. Flow rate is critical for accurate results. Remember to calibrate and recalibrate the pump prior to all sampling. (15 liters/minute)

5. N ever use cassettes that are damaged or expired.

For more information contact Zefon International at www.zefon.com.
Air-O-Cell™ is a registered trademark of Zefon International.

 

Ann Arbor, MI - NVLAP Lab Code 101048-4Atlanta, GA - NVLAP Lab Code 101048-1Baton Rouge, LA - NVLAP Lab Code 200375-0Beltsville, MD - NVLAP Lab Code 200293-0Boston, MA - NVLAP Lab Code 101147-0Buffalo, NY - NVLAP Lab Code 200056-0Calgary, Alberta - NVLAP Lab Code 500100-0Carle Place, NY - NVLAP Lab Code 101048-10Charlotte, NC - NVLAP Lab Code 200841-0Chicago, IL - NVLAP Lab Code 200399-0Corporate - Cinnaminson, NJ - NVLAP Lab Code 101048-0Dallas, TX - NVLAP Lab Code 600111-0Denver, CO - NVLAP Lab Code 200828-0EMSL Canada - Montreal -- Quebec - NVLAP Lab Code 201052-0EMSL Canada -- Toronto - NVLAP Lab Code 200877-0EMSL Canada Inc. -- Ottawa - NVLAP Lab Code 201040-0Fort Lauderdale - NVLAP Lab Code 500085-0Houston, TX - NVLAP Lab Code 102106-0Huntington Beach, CA - NVLAP Lab Code 101384-0Indianapolis, IN - NVLAP Lab Code 200188-0Kernersville, NC - NVLAP Lab Code 102104-0Las Vegas, NV - NVLAP Lab Code 600140-0Miami, FL - NVLAP Lab Code 200204-0Minneapolis IH Lab (EAST) - NVLAP Lab Code 101234-0Minneapolis, MN (WEST) - NVLAP Lab Code 200019-0New York, NY - NVLAP Lab Code 101048-9Orlando, FL - NVLAP Lab Code 101151-0Pasadena, CA - NVLAP Lab Code 200232-0Phoenix, AZ - NVLAP Lab Code 200811-0Piscataway, NJ - NVLAP Lab Code 101048-2Plymouth Meeting, PA - NVLAP Lab Code 200699-0Raleigh, NC - NVLAP Lab Code 200671-0Rochester, NY - NVLAP Lab Code 600183-0S. Portland, ME - NVLAP Lab Code 500094-0Salem, NH - NVLAP Lab Code 201051-0San Diego, CA - NVLAP Lab Code 200855-0San Leandro, CA - NVLAP Lab Code 101048-3Seattle, WA - NVLAP Lab Code 200613-0St. Louis, MO - NVLAP Lab Code 200742-0Vancouver, BC - NVLAP Lab Code 201068-0Wallingford, CT - NVLAP Lab Code 200700-0West Palm Beach, FL - NVLAP Lab Code 600206-0